Evaluation of Diagnostic Radiology Department in Term of Quality Control (QC) of X-Ray Units at Khartoum State Hospitals
No Thumbnail Available
Date
2013
Journal Title
Journal ISSN
Volume Title
Publisher
International Journal of Science and Research
Abstract
The mean objective in diagnostic radiology is to in provide high qualitydiagnostic image while keeping the patients and
workers dose in the lower limit according toALARA principle, (Maria Lucia Nana et al 2009). To optimize the practice of diagnostic
radiology, adequatequality assurance (QA) program should be in place.18 hospitals with 18 x-ray units distributed in Khartoum state
were evaluated inthis study.Each x-ray unit was tested for Kvp andtimeReproducibility , accuracy of Kvp and time , mAs linearity, and
coincidence between light beam and radiation beam. The dark rooms were alsoevaluated to assess the fog level.CONNY II QC
Dosimeter made by PTW company were use to for this study. The analysis of the results showed that two out of eighteenunit had a
problem in mAs linearity, also two out of eighteen unit had a problem in kVp accuracy and one had a problem inkVp reproducibility.
three devices have defects concerning adaptation with optical fieldand radiation field. More than 50% of the darkrooms had a problem
in fog level; time accuracy and time reproducibility were in the acceptable limit.The quality control of the radiologicaldevices should be
performed periodically and regularly and the defects of the devices should be removed in order to beassured of the appropriate function
of the devices. many of these machines need service because of lack ofimplementing the quality control program regularly, which
indicates that the quality control programs should be extendedregularly.because the dark rooms is very important place specially in
conventional radiology departmentlike in Sudan so it need to periodic review to monitor the fog.
Description
Keywords
X-ray, Radiology, quality control, quality assurance